MAGLUMI Laminin (CLIA)
Ref: 130209004M
INTENDED USE
The kit has been designed for the quantitative determination of Laminin (LN) in human serum.
The method can be used for samples over the range of 12.5-1000ng/ml.
The test has to be performed on the Fully-auto chemiluminescence immunoassay (CLIA) analyzer MAGLUMI (Including Maglumi 600,Maglumi 1000,Maglumi 1000 Plus, Maglumi 2000,Maglumi 2000 Plus,Maglumi 3000 and Maglumi 4000).
SUMMARY AND EXPLANATION OF THE TEST
Laminin, a major component of basement membranes, has numerous biological activities including promotion of cell adhesion, migration, growth, and differentiation, including neurite outgrowth. It can be used as a thin coating on tissue-culture surfaces or as a soluble additive to culture medium. Laminin has been shown in culture to stimulate neurite outgrowth, promote cell attachment, chemotaxis, and cell differentation.
For Three Dimensional (3D) culture studies, the High Concentration Laminin/Entactin complex can be used to model the cellular microenvironment in vivo. High concentrations of laminin (2-6 mg/ml) have been used to study acinar differentiation of a human submandibular gland cell line and endothelial cell tubulogenesis.
PRINCIPLE OF THE TEST
Sandwich immunoluminometric assay;
Use an anti-LN monoclonal antibody to label ABEI, and use another monoclonal antibody to label FITC. Sample, Calibrator or Control with ABEI Label, FITC Label and magnetic microbeads coated with anti-FITC are mixed thoroughly and incubated at 37℃, forming a sandwich; After sediment in a magnetic field, decant the supernatant, then cycle washing for 1 time. Subsequently, the starter reagents are added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as RLU within 3 seconds and is proportional to the concentration of LN present in samples.
KIT COMPONENTS
Material Supplies
Reagent Integral for 100 determinations
+ Nano magnetic microbeads: TRIS buffer, 1.2% (W/V), 0.2%NaN3, coated with sheep antiFITC polyclonal antibody: 2.5ml
+ Calibrator Low: bovine serum, 0.2%NaN3: 3.0ml
+ Calibrator High: bovine serum, 0.2%NaN3: 3.0ml
+ FITC Label: anti-LN monoclonal antibody labeled FITC contains BSA, 0.2%NaN3: 6.5ml
+ ABEI Label: anti-LN monoclonal antibody labeled ABEI contains BSA, 0.2%NaN3: 6.5ml
All reagents are provided ready-to-use.
Reagent Vials in kit box
Internal Quality Control: containing BSA, 0.2%NaN3. (target value refer to Quality Control Information date sheet): 2.0ml
Internal quality control is only applicable with MAGLUMI system. Instructions for use and target value refer to Quality Control Information date sheet. User needs to judge results with their own standards and knowledge.
Accessories Required But Not Provided
MAGLUMI Reaction Module REF: 630003
MAGLUMI Starter 1+2 REF: 130299004M
MAGLUMI Wash Concentrate REF: 130299005M
MAGLUMI Light Check REF: 130299006M
Preparation of the Reagent Integral
Before the sealing is removed, gentle and careful horizontal shaking of the Reagent Integral is essential (avoid foam formation!) Remove the sealing and turn the small wheel of the magnetic microbeads compartment to and fro, until the colour of the suspension has changed into brown. Place the Integral into the reagent area and let it stand there for 30 min. During this time, the magnetic microbeads are automatically agitated and completely resuspended.
Do not interchange integral component from different reagents or lots!
Storage and Stability
ï¬ Sealed: Stored at 2-8℃ until the expiry date.
ï¬ Opened: Stable for 4 weeks. To ensure the best kit performance, it is recommended to place opened kits in the refrigerator if it’s not going to be used on board during the next 12 hours.
CALIBRATION AND TRACEABILITY
1)Traceability To perform an accurate calibration, we have provided the test calibrators standardized against the SNIBE internal reference substance.
Calibrators in the Reagent Kit are from Sigma.
2) 2-Point Recalibration
Via the measurement of calibrators, the predefined master curve is adjusted (recalibrated) to a new, instrument-specific measurement level with each calibration.
3) Frequency of Recalibration
ï¬ After each exchange of lot (Reagent Integral or Starter Reagents).
ï¬ Every 2 weeks and/or each time a new Integral is used (recommendation).
ï¬ After each servicing of the Fully-auto chemiluminescence immunoassay (CLIA) analyzer MAGLUMI.
ï¬ If controls are beyond the expected range. ï¬ The room temperature has changed more than 5 ℃ (recommendation)
SPECIMEN COLLECTION AND PREPARATION
Sample material: serum
Collect 5.0ml venous blood into Blood Collection Tube. Standing at room temperature,centrifuging, separating serum part.
The serum sample is stable for up to 12 hours at 2-8℃. More than 12 hours, please packed, -20 ℃ can be stored for 30 days.
Avoid repeated freezing and thawing, the serum sample can be only frozen and thawed two times. Stored samples should be thoroughly mixed prior to use (Vortex mixer).
Please ask local representative of SNIBE for more details if you have any doubt
Vacuum Tubes
(a) Blank tubes are recommended type for collecting samples.
(b) Please ask SNIBE for advice if special additive must be used in sample collecting.
Specimen Conditions
ï¬ Do not use specimens with the following conditions:
(a) heat-inactivated specimens;
(b) Cadaver specimens or body fluids other than human serum;
(c) Obvious microbial contamination.
ï¬ Use caution when handling patient specimens to prevent cross contamination. Use of disposable pipettes or pipette tips is recommended.
ï¬ Inspect all samples for bubbles. Remove bubbles with an applicator stick prior to analysis. Use a new applicator stick for each sample to prevent cross contamination.
ï¬ Serum specimens should be free of fibrin, red blood cells or other particulate matter.
ï¬ Ensure that complete clot formation in serum specimens has taken place prior to centrifugation. Some specimens, especially those from patients receiving anticoagulant or thrombolytic therapy, may exhibit increased clotting time. If the specimen is centrifuged before a complete clot forms, the presence of fibrin may cause erroneous results.
Preparation for Analysis
• Patient specimens with a cloudy or turbid appearance must be centrifuged prior to testing. Following centrifugation, avoid the lipid layer (if present) when pipetting the specimen into a sample cup or secondary tube.
• Specimens must be mixed thoroughly after thawing by low speed vortexing or by gently inverting, and centrifuged prior to use to remove red blood cells or particulate matter to ensure consistency in the results. Multiple freeze-thaw cycles of specimens should be avoided.
• All samples (patient specimens or controls) should be tested within 3 hours of being placed on board the MAGLUMI System. Refer to the SNIBE service for a more detailed discussion of onboard sample storage constraints.
Storage
• If testing will be delayed for more than 8 hours, remove serum from the serum separator, red blood cells or clot. Specimens removed from the separator gel, cells or clot may be stored up to 12 hours at 2-8°C.
• Specimens can be stored up to 30 days frozen at -20°C or colder.
Shipping
Before shipping specimens, it is recommended that specimens be removed from the serum separator, red blood cells or clot. When shipped, specimens must be packaged and labeled in compliance with applicable state, federal and international regulations covering the transport of clinical specimens and infectious substances. Specimens must be shipped frozen (dry ice). Do not exceed the storage time limitations identified in this section of the package insert.
WARNING AND PRECAUTIONS FOR USERS
+ For use in IN-VITRO diagnostic procedures only.
+ Package insert instructions must be carefully followed. Reliability of assay results cannot be guaranteed if there are any deviations from the instructions in this package insert.
Safety Precautions
CAUTION: This product requires the handling of human specimens.
+ The calibrators in this kit are prepared from bovine serum products. However, because no test method can offer complete assurance that HIV, Hepatitis B Virus or other infectious agents are absent; these reagents should be considered a potential biohazard and handled with the same precautions as applied to any serum or plasma specimen.
+ All samples, biological reagents and materials used in the assay must be considered potentially able to transmit infectious agents. They should therefore be disposed of in accordance with the prevailing regulations and guidelines of the agencies holding jurisdiction over the laboratory, and the regulations of each country. Disposable materials must be incinerated; liquid waste must be decontaminated with sodium hypochlorite at a final concentration of 5% for at least half an hour. Any materials to be reused must be autoclaved using an overkill approach. A minimum of one hour at 121℃ is usually considered adequate, though the users must check the effectiveness of their decontamination cycle by initially validating it and routinely using biological indicators.
+ It is recommended that all human sourced materials be considered potentially infectious and handled in accordance with the OSHA Standard on Blood borne Pathogens13. Biosafety Level 214 or other appropriate biosafety practices should be used for materials that contain or are suspected of containing infectious agents.
+ This product contains Sodium Azide; this material and its container must be disposed of in a safe way.
+ Safety data sheets are available on request.
Handling Precautions
ï¬ Do not use reagent kits beyond the expiration date.
ï¬ Do not mix reagents from different reagent kits.
ï¬ Prior to loading the Reagent Kit on the system for the first time, the microbeads requires mixing to re-suspend microbeads that have settled during shipment.
ï¬ For microbeads mixing instructions, refer to the KIT COMPONENTS, Preparation of the Reagent Integral section of this package insert.
ï¬ To avoid contamination, wear clean gloves when operating with a reagent kit and sample.
ï¬ Over time, residual liquids may dry on the kit surface, please pay attention the silicon film still exists on the surface of the kit.
ï¬ For a detailed discussion of handling precautions during system operation, refer to the SNIBE service information.
TEST PROCEDURE
To ensure proper test performance, strictly adhere to the operating instructions of the Fully-auto chemiluminescence immunoassay (CLIA) analyzer MAGLUMI. Each test parameter is identified via a RFID tag on the Reagent Integral. For further information please refer to the Fully-auto chemiluminescence immunoassay (CLIA) analyzer MAGLUMI Operating Instructions.
80μl Sample, calibrator
+40μl ABEI Label
+40μl FITC Label
+20μl Nano magnetic microbeads
15 min Incubation
400μl Cycle washing
3 s Measurement
DILUTION
Sample dilution by analyzer is not available in this reagent kit
Samples with concentrations above the measuring range can be diluted manually. After manual dilution, multiply the result by the dilution factor.
Please choose applicable diluents or ask SNIBE for advice before manual dilution must be processed.
QUALITY CONTROL
+ Observe quality control guidelines for medical laboratories
+ Use suitable controls for in-house quality control. Controls should be run at least once every 24 hours when the test is in use, once per reagent kit and after every calibration. The control intervals should be adapted to each laboratory’s individual requirements. Values obtained should fall within the defined ranges. Each laboratory should establish guidelines for corrective measures to be taken if values fall outside the range.
LIMITATIONS OF THE PROCEDURE
1) Limitations
A skillful technique and strict adherence to the instructions are necessary to obtain reliable results. Procedural directions must be followed exactly and careful technique must be used to obtain valid results. Any modification of the procedure is likely to alter the results.
Bacterial contamination or repeated freeze-thaw cycles may affect the test results
2) Interfering Substances
No interference with test results is seen by concentrations of bilirubin<0.06mg/ml, haemoglobin<16mg/dl or triglycerides< 12.5mg/ml.
3) HAMA
Patient samples containing human anti-mouse antibodies (HAMA) may give falsely elevated or decreased values. Although HAMA-neutralizing agents are added, extremely high HAMA serum concentrations may occasionally influence results.
4) High-Dose Hook
No high-dose hook effect was seen for LN concentrations up to 10μg/ml.
RESULTS
1) Calculation of Results
+ The analyzer automatically calculates the Laminin concentration in each sample by means of a calibration curve which is generated by a 2-point calibration master curve procedure. The results are expressed in ng/ml. For further information please refer to the Fully-auto chemiluminescence immunoassay (CLIA) analyzer MAGLUMI Operating Instructions.
2) Interpretation of Results
ï¬ Results of study in clinical centers with group of individuals, 95% of the results were:< 50 ng/ml.
ï¬ Results may differ between laboratories due to variations in population and test method. If necessary, each laboratory should establish its own reference range.
PERFORMANCE CHARACTERISTICS
1) Precision
Intra-assay coefficient of variation was evaluated on 3 different levels of control serum repeatedly measured 20 times in the same run, calculating the coefficient of variation.
Intra-assay precision
Control Mean(ng/ml) SD(ng/ml) CV%
Level 1 121.35 6.61 5.45
Level 2 182.36 9.37 5.14
Level 3 658.93 31.96 4.85
Inter-assay coefficient of variation was evaluated on three batches of kit. Repeatedly measured 3 different levels of control serum 21 times, calculating the coefficient of variation.
Inter-assay precision
Control Mean(ng/ml) SD(ng/ml) CV%
Level 1 118.46 10.48 8.85
Level 2 192.57 16.43 8.53
Level 3 672.45 54.94 8.17
2) Analytical Sensitivity
The sensitivity is defined as the concentration of Laminin equivalent to the mean RLU of 20 replicates of the zero standard plus two standard deviations corresponding to the concentration from the standard curve. The sensitivity is typically less than 12.5ng/ml.
3) Specificity
The specificity of the Laminin assay system was assessed by
measuring the apparent response of the assay to various
potentially cross reactive analytes.
Compound Concentration Cross reactivity
CI 200 ng/ml 6.25%
CII 500 ng/ml 2.5%
CIII 800 ng/ml 1.6%
CIV 1000 ng/ml 1.25%
4) Recovery
Consider calibrator high of known concentration as a sample,
dilute it by 1:2 ratio with diluents, and measure its diluted
concentration for 10 times. Then calculate the recovery of
measured concentration and expected concentration. The
recovery should be within 90% -110%
Expected Mean Measuring Recovery
343.8 ng/ml 351.2 ng/ml 102%
5) Linearity
Use Laminin calibrator to prepare the six-point standard curve,
measuring all points’ RLU except point A, and then do four
parameter linear fitting in double logarithm coordinate, the
absolute linear correlation coefficient(r) should be bigger than
0.9800.
Calibrator
Point Concentration
ng/ml Absolute linear
correlation coefficient (r)
A 0.0
B 50.0 r=0.9970
C 100.0
D 200.0
E 500.0
F 1000.0
6) Method comparison
A comparison of MAGLUMI Laminin (y) with a commercially
available Laminin test (x) using clinical samples gave the following
correlations (ng/ml):
Linear regression
y = 0.3819x+2.6652
r = 0.9470
Number of samples measured: 100
The sample concentrations were between 22.45 and 103.12
ng/ml.
REFERENCES
1.Hewitt RE,Powe DG,Morrell K,et al.Laminin and Collagen â…£
subunit distribution in normal and neoplastic tissues of
colorectum and breast.Br J Cancer, 1997, 75(2):221.
2.Orre M, Rogers PA. Reduced vascular
basement-membraneimmunostaining in mucinous tumors of
the ovary.Int J Cancer, 1998, 79(2):139.
3.Enam SA,Rosenblum ML,Edvardsen K,Role of extracellular matrix in tumor invasion:migration of glioma cells along
fibronectin-positive mesenchymal cell processes. Neurosurgery,
1998, 42(3):599.
4.Ambiru S,Miyazaki M,Nakagawa K,et al.Increased serum type
â…£ collagen 7-S levels in patients with hepatic etastasis.Am J
Gastroenterology, 1995, 90(5):783.
5.Khan A,Krishna M,Baker SP,et al.Cathepsin B expression and
its correlation with tumorassociated laminin and tumor
progression in gastric cancer.Arch Pathol Lab
Med,1998,122(2):172.
6.Orita H,Korenaga D,Maehara Y et al.Laminin distribution
patterns are closely related to liver metastasis in gastriv
cancer.Cancer,1993,71(4):1201.
7.Pares A, Deulofeu R, Gimenez A, Caballeria L, Bruguera M,
Caballeria J, Ballesta AM, Rodes J: Serum hyaluronate reflects
hepatic fibrogenesis in alcoholic liver disease and is useful as a
marker of fibrosis. Hepatology 1996, 24(6): 1399-1403.
8.Oberti F, Valsesia E, Pilette C, Rousselet MC, Bedossa P, Aube
C, Gal-lois Y, Rifflet H, Maiga MY, Penneau-Fontbonne D,
Cales P: Nonin-vasive diagnosis of hepatic fibrosis or cirrhosis.
Gastroenterology 1997, 113(5):1609-1616.
9.Wai CT, Greenson JK, Fontana RJ, Kalbfleisch JD, Marrero JA,
Con-jeevaram HS, Lok AS: A simple noninvasive index can
predict both significant fibrosis and cirrhosis in patients with
chronic hepatitis C. Hepatology 2003, 38(2): 518-526.